INTRODUCTION: Today, the increasing number of incidentally detected peripheral pulmonary lesions (PPLs) within and outside lung cancer screening trials is a diagnostic challenge. This fact encourages further improvement of diagnostic procedures to increase the diagnostic yield of transbronchial biopsy, which has been shown to have a low complication rate. The purpose of this study was to evaluate the safety and feasibility of a new ultrathin 1.1 cryoprobe that can be placed through an ultrathin bronchoscope (UTB) using fluoroscopy and radial endobronchial ultrasonography (rEBUS) navigation for assessing PPLs. METHODS: Thirty-five patients with PPL less than 4 cm in diameter were prospectively enrolled to receive transbronchial cryobiopsies (TBCBs) using the ultrathin 1.1-mm cryoprobe. Navigation to the PPL was accomplished with the UTB. Under rEBUS and fluoroscopy guidance up to 4 cryobiopsies were obtained. The sample sizes of the biopsies were compared to a historic collective derived from a 1.9-mm cryoprobe and standard forceps. The feasibility and safety of the procedure, the cumulative and overall diagnostic yield, and the cryobiopsy sizes were evaluated. RESULTS: After detection with the rEBUS, TBCB was collected from 35 PPLs, establishing a diagnosis in 25 cases, corresponding to an overall diagnostic yield of 71.4%. There was no difference in diagnostic yield for PPL <20 mm or
Keywords